技術(shù)參數(shù)
GOVISC SPECIFICATION | ||
PERFORMANCE | Range | 0 – 100,000cP over 5 models |
Measurement Time | 1 second | |
Accuracy | 1% of reading or +/- 1 digit | |
Repeatability | 0.3% of reading or +/- 1 digit | |
OUTPUTS | Analog | 4-20mA configurable, representing viscosity, ref. viscosity, or temperature |
Digital | RS485 ModBus Data Link to host system or PC running ViscoLink | |
USB | USB II Plug and Play through VP100 Interface | |
TEMPERATURE | Integral Thermometer | Integrated tip mounted temperature measurement |
Standard Range | -20 °C to 150 °C | |
Extended Range | To order for greater than 150 °C * | |
SENSOR PROJECTION | Nominal sensor projection | 300mm |
ELECTRICAL CONNECTION | Cable Length | As required – not affected by cable length. |
Cable/Connector Type | Tyco M9 | |
Power | USB powered through VP100 Interface | |
CONSTRUCTION | All welded, crevice-free | |
MATERIALS | Sensor | SS316 ( EN 1.4401) |
Sensor Options: | Alloy C22 (EN 2.4602) | |
Alloy C276 (EN 2.4819) | ||
Special Materials: | Non-standard materials to order * | |
SURFACE FINISH | Standard | N5 0.4 Ra |
Finish Options | Fluoropolymer | |
Electropolished | ||
Electroplated | ||
Special Finish | Non-standard finish options and coatings to order * | |
WEIGHT | 700g | |
INGRESS PROTECTION | Standard | IP65 |
SOFTWARE | ViscoLink data viewing and logging PC software | |
* SPECIAL OPERATIONS | Custom variations to instrument specification and dimensions routinely provided by our dedicated Special Operations division | |
As it is Hydramotion’s policy to continuously develop and improve its products, some specifications may change without notice |