Steady-State Food & Drug Stability Chambers |
Blue M’s steady-state chambers are available in four convenient sizes. With workspace sizes of 8, 16, 32, and 58-cubic-feet and a variety of temperature and humidity combinations, these chambers provide the flexibility for a wide range of uses.
Watlow F4 Controller allows remote and unattended operation in setting up programs and setpoints. The F4 also allows easy control over both temperature and humidity settings in one controller. Common design elements of the chambers allow shelves to be interchanged among different models. Streamlined design allows Thermal Product Solutions to ship chambers customized to the customer’s needs in far less time than previous product lines. Customers with several CEO chambers, even if different sizes, now have consistent controls, utility requirements, and heat output, so plant management issues are greatly reduced.
Stability and Shelf-Life Chambers Features: |
Remote and unattended operation in setting up programs and setpoints |
Easy control over both temperature and humidity settings |
Non-ramping, manual mode |
Serial communications capabilities via the EIA-232 and EIA-485 standards |
Modbus RTU protocol |
Direct printing to a serial printer |
Direct connection to a PLC, or computer workstation |
High-definition LCD display with four digital input lines |
Two alarm outputs and eight digital event outputs |
Installation has also been improved with the addition of new combination casters/leveling legs |
Stability and Shelf-Life Chambers Specifications: |
Model | Non-humidified | CEO 908-2 | CEO 916-2 | CEO 932-2 | CEO 958-2 | |||||
Humidified | CEO 908-4 | CEO 916-4 | CEO 932-4 | CEO 958-4 | ||||||
Workspace inches/cm | W | 33/84 | 33/84 | 33/84 | 60/152 | |||||
D | 28/71 | 28/71 | 28/71 | 28/71 | ||||||
H | 15/38 | 30/76 | 60/152 | 60/152 | ||||||
Exterior inches/ cm | W | 39/99 | 39/99 | 39/99 | 70/178 | |||||
D | 32/81 | 32/81 | 32/81 | 32/81 | ||||||
H | 46/117 | 61/155 | 91/231 | 91/231 | ||||||
No. of Shelves | 1 | 2 | 4 | 8 | ||||||
No. of Shelf Guides | 4 | 9 | 19 | 38 | ||||||
Temperature Range | Low | 0 | ||||||||
High | 99°C | |||||||||
Utilities | Refrigeration (HP) | 2-Jan | 4-Mar | |||||||
Heater Capacity | 2 kW | 6 kW | ||||||||
Humidifier | kW | 1.5 | 3 | |||||||
GPH | 0.5 | 1 | ||||||||
AMPS @ 208/230, 3-WIRE, 1Ph-60Hz | 20 | 24 | 20 | 24 | 20 | 24 | 43 | 56 | ||
AMPS Breaker | 30 | 30 | 30 | 30 | 30 | 30 | 60 | 70 | ||
Over-current protection recommended | ||||||||||
Unit Weight | Lbs. | 375 | 400 | 425 | 450 | 524 | 550 | 1850 | 1900 |