3-2839 / 3-2840 / 3-2841 / 3-2842
+GF+ SIGNET2839-2842 電導(dǎo)率/ 電阻率電極,可以在0.01-10.0 cm-1四個單元常數(shù)內(nèi)應(yīng)用。并適合廣泛的各種應(yīng)用范圍,從高純水質(zhì)量監(jiān)測到除鹽還原裝置監(jiān)測。316SS的電極表面經(jīng)過精密的噴砂作業(yè),控制了光潔度,從而確保了測量的**度和可重復(fù)性。PEEK?絕緣體和過程連接器是整體注塑的,使電極間的差異降低到*小值。雙螺紋連接器,不是3/4 in.NPT螺紋就是ISO7/1-R3/4 in.螺紋,確保浸入式和在線式結(jié)構(gòu)的安裝中能快速而方便。變送器的一體化安裝套件和接線盒是可以采用的附件。
0.055 to 100μS (18.3MΩ to 10KΩ)
(0.02 to 50 ppm)
3-2840測量范圍:0.1 cm-1
1 to 1,000μS (1MΩ to 1KΩ)
(0.5 to 500 ppm)
3-2841測量范圍:1.0 cm-1
10 to 10,000μS (5 to 5,000 ppm)
3-2842測量范圍:10.0 cm-1
100 to 200,000μS (50 to 100,000 ppm)
可重復(fù)性:±2%滿量程.
精度:±1%滿量程
2839-2842電導(dǎo)率探頭詳細英文介紹
2839-2842 NPT Conductivity-Resistivity Electrodes
The 2839-2842 Conductivity/ Resistivity Electrodes are available in four cell constants from 0.01 to 10.0 cm-1, and are suitable for a wide variety of applications from high purity water quality monitoring to weak acids and bases. 316SS electrode surface finishes are controlled in a precision bead blasting operation to ensure measurement accuracy and repeatability.
The PEEK™ insulator and process connections are injection over-molded to minimize variance between electrodes. Double threaded connections in either 3/4 in. NPT or ISO 7/1-R 3/4 enable quick and easy installation in submersible or in-line configurations. Transmitter integral mounting kit and junction boxes are available as accessories.
Online Catalog
Fields of Application:
- Water Treatment & Water Quality Monitoring
- Reverse Osmosis
- Deionisation
- Cooling Tower and Boiler Protection
- Distillation
- Desalination
- Demineraliser
- Semiconductor
- Aquatic Animal Life Support Systems
Advantages:
- Controlled surface finish ensures accuracy and repeatability
- 316SS electrodes with injection molded PEEK™ process connections and insulators
- Compact electrode length for easy in-line installation
- Suitable for operation up to 131 °C (268 °F)
- 3/4 in. NPT or ISO 7/1 - R 3/4 process connections
- Double-threaded connection for in-line or submersible mount
- Triple orifice flow-through design reduces clogging and bubble entrapment
- Cell constants may be traceable to NIST and certified to within ±1% of value
- Integral Mounting Kit and junction box acccessories
Standards:
- RoHS compliant
Downloads (選型手冊下載) Data Sheet - 2839-2842 Conductivity Electrodes (pdf, 0.5 MB) |
Manuals (使用說明書下載) English (pdf, 0.3 MB) |
Compatible Products (兼容產(chǎn)品鏈接) 5800CR ProPoint Conductivity/Resistivity Monitor 5900 ProPoint Salinity Monitor 8850 ProcessPro Conductivity/Resistivity Transmitter 8860 ProcessPro Dual Channel Conductivity/Resistivity Transmitter |