DM4DL超聲波測(cè)厚儀
Krautkramer超聲波測(cè)厚儀DM4系列中的DM4DL,小巧輕便,操作簡(jiǎn)單,功能。DM4DL*的DUAL-MULTI模式--即測(cè)量穿過(guò)涂層的操作模式,在被測(cè)物表面有油漆時(shí),無(wú)需去除油漆而測(cè)量材料的凈厚度(鋼管\壓力管道\鍋爐壁厚的測(cè)量)。
DM4DL的特點(diǎn):
- 可配標(biāo)準(zhǔn)、超厚、超薄和高溫探頭
- 自動(dòng)探頭識(shí)別、V聲程校正、零位校正
- 帶zui小讀數(shù)顯示和存儲(chǔ)的zui小測(cè)量模式
- 穿過(guò)涂層測(cè)量厚度的操作模式
- 在不同模式的壁厚測(cè)量
- 極值LED報(bào)警
- 數(shù)據(jù)記錄儀可存儲(chǔ)5390個(gè)數(shù)據(jù)
- RS232接口可接打印機(jī)和計(jì)算機(jī)
DM4DL的技術(shù)參數(shù)
測(cè)量范圍 |
0.6~300mm(鐵) |
DM4DL的常用探頭技術(shù)數(shù)據(jù)
名稱(chēng) | 型號(hào) | 測(cè)量范圍 | 頻率 | 探頭接觸直徑 | 工作溫度 | 測(cè)量zui小鋼管直徑 | 探頭線型號(hào) |
標(biāo)準(zhǔn)探頭 | DA301 | 1.2~200mm | 5MHz | 12.1mm | -20~60℃ | 25mm | DA231 |
超厚探頭 | DA303 | 5.0~300mm | 2MHz | 16.2mm | -20~60℃ | 127mm | DA231 |
超薄探頭 | DA312 | 0.6~50mm | 10MHz | 7.5mm | -20~60℃ | 20mm | DA235 |
超薄探頭 | DA312B16 | 0.7~12mm | 10MHz | 3.0mm | -20~60℃ | 15mm | 自帶導(dǎo)線 |
高溫探頭 | DA305 | 4.0~60mm | 5MHz | 16.2mm | 10~600℃ | DA235 | |
高溫探頭 | HT400 | 1.0~254mm | 5MHz | 12.7mm | 10~540℃ | 50mm | KBA535 |
注:所示溫度為廠家理想狀態(tài)試驗(yàn)所得,實(shí)際溫度會(huì)低于此溫度。注意高溫耦合劑的正確使用方法。
測(cè)試高溫表面時(shí)需用高溫耦合劑。
DM4DL的標(biāo)準(zhǔn)配置
主機(jī),DA301探頭,DA231導(dǎo)線,耦合劑,兩節(jié)5號(hào)電池,螺絲刀,儀器軟包,中英文說(shuō)明書(shū),儀器箱
ccccccccccccc
DM4, DM4DL Thickness Gauge
The DM4 series enables you to accomplish a large number of demanding tasks, especially with remaining wall thickness measurements on components subjected to corrosionerosion.
The DM4DM4DL are smart, simple-to-use ultrasonic thickness gauges that offer big ultrasonic performance in a small lightweight package. Compact sizeergonomic shape provide comfortable one-hand use.
DM4, DM4DL Thickness Gauge Products Features
- Equipped with a membrane keypadan impact-resistant, dustproofwatertight housing to guarantee long-lastingreliable operation
- Field-proven features such as Minimum Capture Mode, Differential Mode2-Point Calibration
- Variety of application for the most different materials is made possible by the large selection of probes including inligent dialog probes for optimized settingimproved performance.
- Automatic probe zeroing for optimized gauge setting to a large number of probestest conditions
- DUAL MULTI operating mode for accurate wall thickness measurement on painted or coated pipelinestanks. Any time-consumingcost-intensive removal of the coating is no longer needed
- Data logger for storing up to 5,000 thickness readings for later download to software for viewing, printing reportsdigitally storing data.
- UltraMATE software for data management (DM4-DL only)
DM4, DM4DL Thickness Gauge Applications
The DM4 instrument series are suitable for numerous test tasks on components subject to wear:
- Pipelines, pressure vessels, gas cylinders, storage tanks, vessels for chemical processes
- Materials handling systems, pumps
- Bridges, hulls, cranes, equipment for shipbuildingstructural steel engineering
High temperature probes for temperatures up to 600°C enables you to perform thickness measurements on extremely hot components.